CLab-LT FAQs

MGICLab-LT

CLab-LT - Frequently Asked Questions

How does the sample library maintain thermal insulation?

The sample library uses insulated panels with an outer color steel layer, an inner stainless steel layer, and polyurethane insulation cotton in between. The base also uses a double-insulation structure to help maintain stable low-temperature storage.

The sample library is operated by authorized personnel through the software interface. Before sample access, an application must be submitted in the system and approved by relevant personnel to generate an order.

The storage temperature is -80±5°C.

The storage capacity is customized according to the site and can range from 500,000 to several million samples.

Storage capacity is calculated using SBS 0.75 mL cryovials as the conversion reference, following international industry practice.

The system binds sample information by scanning the 1D code of the sample box and the 2D code of the sample tube during storage and retrieval. If a sample does not match the stored information, it will be rejected.

It supports multiple specifications of SBS cryovials from mainstream brands, including 0.5 mL, 0.75 mL, 1 mL, 2 mL, and 5 mL formats.

If only whole-box storage and retrieval are needed, EP tubes can be used. However, they do not support tube picking, so single-tube storage and retrieval is not possible.

The room should not exceed 25°C in temperature and 50% RH in humidity. It should also have ventilation or fresh air, no obvious vibration sources, and sufficient power, lighting, and drainage preparation.

The system includes UPS backup power, which supports continued operation of the automated structure for 0.5 hours to complete the storage or retrieval process. After a power failure, liquid nitrogen supply helps maintain the internal temperature.