GenSIRO-48

The GenSIRO-48 Fully Automated Sequencing Library Preparation System integrates sample processing, nucleic acid extraction, library preparation, and quality control on a single automated platform. Supporting flexible batches of 1–48 samples, it automates the entire workflow from sample loading to sequencing library output, reducing manual intervention and streamlining sequencing workflows.

Physically separated extraction and library preparation compartments, automated sample transfer, HEPA filtration, and UV disinfection work together to minimize cross-contamination risks. With full-process traceability and standardized workflows, GenSIRO-48 delivers efficient, reliable library preparation across diverse applications. The GenSIRO-48 series is CE-marked, with RUO and IVD versions available to support both research and clinical needs.

Product Overview

GenSIRO-48 is MGI's fully automated sequencing library preparation system. Process 1–48 samples from primary tubes to sequencing-ready libraries.
GenSIRO-48 is MGI's fully automated sequencing library preparation system. Process 1–48 samples from primary tubes to sequencing-ready libraries.

Performance Parameters

Sample Types & Loading Capacity
Sample Capacity 1–48 tubes, with automated cap removal
Compatible Sample Tubes Blood collection tubes, centrifuge tubes, cryogenic tubes, EP tubes
Liquid Handling Module
Pipetting Range 10–1,000 µL (8-channel, 1 mL)
2–200 µL (8-channel, 200 µL)
1–1,000 µL (single-channel, 1 mL)
8-Channel 1 mL Pipettor 10 µL: CV ≤5%, Accuracy ±5%
100 µL: CV ≤1%, Accuracy ±2%
200 µL: CV ≤1%, Accuracy ±2%
500 µL: CV ≤0.5%, Accuracy ±1%
1,000 µL: CV ≤0.5%, Accuracy ±1%
8-Channel 200 µL Pipettor 2 µL: CV ≤5%, Accuracy ±10%
10 µL: CV ≤2%, Accuracy ±5%
50 µL: CV ≤1%, Accuracy ±2%
200 µL: CV ≤1%, Accuracy ±1%
Single-Channel 1 mL Pipettor 1 µL: CV ≤8%, Accuracy ±10%
2 µL: CV ≤8%, Accuracy ±10%
100 µL: CV ≤1%, Accuracy ±2%
500 µL: CV ≤0.5%, Accuracy ±1%
1,000 µL: CV ≤0.5%, Accuracy ±1%
Consumable Compatibility
Compatible Plates U-/V-bottom 96-well deep-well plates, full-skirt PCR plates
Temperature Control Module
Temperature Range 4–90°C
Temperature Accuracy (6°C, 37°C, 42°C, 55°C) ±1°C
Temperature Uniformity (6°C, 37°C, 42°C, 55°C) ≤1.5°C
Temperature-Controlled Shaking Module
Temperature Range Ambient to 90°C
Temperature Accuracy (55°C) ±1°C
Temperature Uniformity (55°C) ≤2°C
Shaking Speed Range 100–2,000 rpm
Shaking Speed Stability Within ±30 rpm
Temperature Range 4–99°C
Temperature Control Precision (55°C, 72°C, 95°C) ≤0.5°C
Temperature Accuracy (55°C, 72°C, 95°C) ±0.5°C
Temperature Uniformity (55°C, 72°C, 95°C) ≤1°C

System Parameters

Power Requirements, Dimensions & Net Weight
Input Voltage 200–240 V AC, 50/60 Hz
Rated Power 3,000 VA (Input 1: 1,000 VA; Input 2: 2,000 VA)
Dimensions (W × D × H) 2,200 × 870 × 1,900 mm (excluding the height required to open the viewing window)
Net Weight Approx. 750 kg
Floor Load Capacity >500 kg/m²
Operating Environment
Temperature Range 19–25°C
Relative Humidity 20–70% RH, non-condensing
Atmospheric Pressure 80–106 kPa
Maximum Operating Altitude 2,000 m
Pollution Degree 2
For indoor use only.
Storage & Transportation Conditions
Temperature Range −20 to 50°C
Relative Humidity 15–85% RH, non-condensing
Atmospheric Pressure 80–106 kPa
GenSIRO-48 is MGI's fully automated sequencing library preparation system. Process 1–48 samples from primary tubes to sequencing-ready libraries.

Features

End-to-End Automation & Traceability

  • From sample loading to sequencing-ready libraries in just two simple steps, with automated barcode scanning and data collection for comprehensive sample tracking and workflow traceability.

Flexible Sample Processing

  • Supports 1–48 primary sample tubes per run and accommodates blood, body fluids, tissues, cells and swabs without the need for batch accumulation.

All-in-One Integration

  • Integrates extraction, liquid handling, PCR and library preparation into a compact footprint of less than 2 m².

Advanced Contamination Control

  • A dual-compartment design with differential pressure, HEPA filtration and UV disinfection minimises cross-contamination risks.

Discover More

Get in Touch

Email
MGI-service@mgi-tech.com

Phone
📞(+86) 4000-688-114

Marketing & PR
MGI_Branding@mgi-tech.com

Careers
MGI_Recruiting@mgi-tech.com

Investor Relations
MGI_IR@mgi-tech.com

Message us